Rumi Adapter Cleaning and Sterilizing Instructions
Cleaning
-
Remove the inner rod from the outer arm.
-
Disconnect the Rumi Adapter from the from the inner rod
assembly at the ball joint and remove the joing screw from
the Rumi Adapter. (See pg. 7 of the Rumi Instruction Manual)
-
Inspect parts for defects, such as parts not fitting together
or misshapen parts. Contact Side-Kick sales reps. regarding
defects.
-
Use Enzol® enzymatic detergent, and prepare according to
the manufacture's recommendations of 1 oz. per gallon of warm
tap water, or an FDA approved equivalent.
-
Soak the outer arm, inner rod and Rumi Adapter for five
minutes in the prepared solution.
-
Maneuver movable parts to loosen trapped soil. Use a soft
bristled brush (M16) to gently scrub to remove any visible
soil and actuate all moveable parts while scrubbing in order
to loosen trapped soil. Use an appropriate size lumen brush
(45-541) for internal channels.
-
Repeat steps 5 and 6 two addional times for a total of three
times.
-
Prepare a tap water bath at 38-49℃.
-
Agitate the assembly by hand for a minimum of one minute in
the prepared bath.
-
Rinse the assembly under running tap water for at least one
minute.
-
Dry the exterior of the assembly with a clean, lint free cloth.
Sterilizing
-
Place items in the Kronner Sterilizer tray.
-
Sterilization wrap: Kimgard KC400 or FDA approved equivalent
-
Pre-vacuum temperature: 270℉ (132℃)
-
Exposure time: 4 minutes
-
Drying time: 45 minutes