Kronner Medical

Universal Adapter Cleaning and Sterilizing Instructions


Cleaning

  1. Remove the connection screw.
  2. Use Enzol® enzymatic detergent and prepare according to the manufacture's recommendations of 1 oz. per gallon of warm tap water.
  3. Soak the adapter and connection screw for five minutes in the prepared solution.
  4. Maneuver parts to loosen trapped soil. Use a soft bristled brush (M16) to gently scrub to remove any visible soil.
  5. Repeat Step 4 two additional times for a total of three times.
  6. Prepare a tap water bath at 38-49℃.
  7. Agitate the components by hand for a minimun of one minute in the prepared bath.
  8. Rinse the assembly under running tap water for at least one minute.
  9. Dry the exterior of the assembly with a clean, lint-free cloth.

Sterilizing

  1. Place items in a sterilizer tray.
  2. Sterilization wrap: Kimgard KC400
  3. Pre-vacuum temperature: 270℉ (132℃)
  4. Exposure time: 4 minutes
  5. Drying time: 45 minutes