Universal Adapter Cleaning and Sterilizing Instructions
Cleaning
-
Remove the connection screw.
-
Use Enzol® enzymatic detergent and prepare
according to the manufacture's recommendations
of 1 oz. per gallon of warm tap water.
-
Soak the adapter and connection screw for five
minutes in the prepared solution.
-
Maneuver parts to loosen trapped soil. Use a
soft bristled brush (M16) to gently scrub to
remove any visible soil.
-
Repeat Step 4 two additional times for a total
of three times.
-
Prepare a tap water bath at 38-49℃.
-
Agitate the components by hand for a minimun of
one minute in the prepared bath.
-
Rinse the assembly under running tap water for
at least one minute.
-
Dry the exterior of the assembly with a clean,
lint-free cloth.
Sterilizing
-
Place items in a sterilizer tray.
-
Sterilization wrap: Kimgard KC400
-
Pre-vacuum temperature: 270℉ (132℃)
-
Exposure time: 4 minutes
-
Drying time: 45 minutes