VCM / Vcare Connector Cleaning and Sterilizing Instructions
Cleaning
- Remove the connection screw.
- Use Enzol® enzymatic detergent and prepare according to the manufacture's recommendations of 1 oz. per gallon of warm tap water.
- Soak the adapter and connection screw for five minutes in the prepared solution.
- Maneuver parts to loosen trapped soil. Use a soft bristled brush (M16) to gently scrub to remove any visible soil.
- Repeat Step 4 two additional times for a total of three times.
- Prepare a tap water bath at 38-49℃.
- Agitate the components by hand for a minimun of one minute in the prepared bath.
- Rinse the assembly under running tap water for at least one minute.
- Dry the exterior of the assembly with a clean, lint-free cloth.
Sterilizing
- Place items in a sterilizer tray.
- Sterilization wrap: Kimgard KC400
- Pre-vacuum temperature: 270℉ (132℃)
- Exposure time: 4 minutes
- Drying time: 45 minutes