VCR Cleaning and Sterilizing Instructions
Cleaning
-
Detach the Universal or Vcare Adapter from the
Side-Kick Arm.
-
Remove the joing screw from the chosen adapter
and connector screw from the Radel Connector.
-
Remove the Vaginal Cuff Retractor, Connector,
chosen adapter, inner rod and outer arm.
-
Use Enzol® enzymatic detergent and prepare
according to the manufacture's recommendations
of 1 oz. per gallon of warm tap water, or use
an FDA approved equivalent.
-
Soak all the parts for five minutes in the
prepared solution.
-
Maneuver parts to loosen trapped soil. Use a
soft bristled brush (M16) to gently scrub to
remove any visible soil.
-
Repeat step 6 two additional times for a total
of three times.
-
Prepare a tap water bath at 38-49℃.
-
Agitate the components by hand for a minimum of
one minute in the prepared bath.
-
Rinse the components under running tap water
for at least one minute.
-
Dry the components with a clean, lint free cloth.
Sterilizing
-
Place inner rod, outer arm, adapters, connector
and screws in a sterilizer tray.
-
Sterilization wrap: Kimgard KC400
-
Pre-vacuum temperature: 200℉ (132℃)
-
Exposure time: 4 minutes
-
Drying time: 45 minutes